| Country of Origin | UK |
|---|---|
| Repeatability | RSD ≤ 1% |
| Brand | Labtron |
| Power Supply | 220 ± 22 VAC / 50 ± 1 Hz |
| Type | All-solid-state RF generator |
| Net Weight | 192.4 Kg |
| Working Temperature | 15°C to 30°C |
| Model | LICP-C10 |
| Wavelength Range | 165 to 950 nm |
| Stability | RSD ≤ 1.5% @ 2 hrs |
| Stray Light | Equivalent background concentration Less than 2 µg/ml at 189.042 nm |
| Gross Weight | 249.5 Kg |
| Detector Type | Large-area CCD / full-spectrum response / high UV quantization efficiency |
| Temperature Control | 38 ± 0.1°C |
| Type of Product | Spectrometer |
| Sampling Systems | Optimized for organic solvents / high-salt/complex matrix samples |
| Atmospheric pressure | 86 to 106 KPa |
| WORKING HUMIDITY | ≤ 70% |
| Storage and transportation temperature | 5°C to 40°C |
| Storage Relative Humidity | ≤ 85% |
| Net Dimension | 1270 x 720 x 850 mm |
| Grating Specifications | Echelle grating / 100 x 50 mm |
| Observation Design | Radial and axial observation for element measurement from sub-ppm to high content |
| Optical Path Structure | Echelle grating and prism cross-dispersion with ultra-pure SiO2 prism |
| Imaging Quality | Optimized design with aspherical optical elements |
| Light Chamber Technology | Multi-point filling for faster operation |
| Gas Path | Independent gas path for nitrogen or argon |
| Spectroscopic System | Medium-step 2D optical system with 400 mm focal length |
| Detector Pixel | 1024 x 1024 pixels / 24 x 24 µm pixel area |
| Detector Working Temperature | Less than -40°C / stabilization time Less than 3 minutes |
| Detector Quantization Efficiency | Greater than 75% without coating |
| RF Generator Cold Cone Tail Flame Technology | Minimizes self-absorption and ionization interference |
| RF Generator Power Output | 700W to 1600W / continuously adjustable by 1W |
| RF Generator Power Stability | ≤ 0.1% |
| RF Generator Oscillation Frequency | 27.12 MHz |
| RF Generator Frequency Stability | ≤ 0.01% |
| RF Generator Electromagnetic Field Leakage | Less than 0.5 V/m |
| Feedback Diverter System | Two-inlet and six-out waterway for better control |
| Peristaltic Pump | 4-channel / 12-roller with adjustable speed |
| Argon Gas Consumption | 8 L/min to 18 L/min |
| Cooling Gas | 0.00 L/min to 20.00 L/min / accuracy 0.01 L/min |
| Auxiliary and Carrier Gas | 0.00 L/min to 2.00 L/min / accuracy 0.01 L/min |
| Multi-Method Analysis | Supports simultaneous measurement / editing / and viewing |
| Spectral Line Library | Greater than 70 /000 spectral lines / intelligent interference prompts |
| Method Support | Standard curve and standard addition methods / multiple data processing methods |
| Report Output | Multiple formats available |
| Liquid Content Range | 0.01 ppm to several thousand ppm |
| Solid Content Range | 0.001% to 70% |
| Element Detection Limit | 1 to 10 ppb for most elements |
| Gross Dimension | 1500 x 900 x 1060 mm |
ICP Optical Emission Spectrometer LICP-C10 operates over a wavelength range of 165 to 950 nm. It features a dual observation design, allowing for both radial and axial measurements. The sampling systems are optimized for organic solvents, high-salt, and complex matrix samples, ensuring accurate analysis in challenging conditions. A mass flow controller of our ICP optical emission spectrometer regulates the flow of cooling, auxiliary, and carrier gases to ensure optimal performance and efficiency in various applications.
Applications:
ICP Optical Emission Spectrometer LICP-C10 is used to analyze trace elements in substances soluble in nitric, hydrochloric, and hydrofluoric acids, in fields like environmental protection, petroleum, rare earths, semiconductors, geology, metallurgy, clinical medicine, food safety, forensics, and agriculture.